In the event of unexplained, inconsistent weight measurements:

  1. Make sure the weighing lysimeter connector is properly connected.
    Unplug it with a small, thin, flat head screwdriver (b) pushed between the plug and the socket and pull gently. Be careful not to break the connectors by pushing too hard.
  2. Check the weighing lysimeter connector for wiring problems, physical damage, or excessive moisture. 
  3. Reboot the controller:
    For Version 2.0: Unplug the incoming power cable for 5 seconds and plug it back in.
    For Version 3.0: In the "Control Panel" section of Plantarray Client, click "Reset PIU ".
  4. Make sure that the calibration values did not change for any reason.
  5. If the problem persists, please contact us.